The wall-mounted KS-500B intelligent multi parameter analyzer cabinet can synchronously and correctly detect pH, ORP, residual chlorine, turbidity, and temperature. Data capture, display, storage, dosage control, and remote transmission are integrated. Easy to use, precise measurement, wireless remote monitoring, over limit alarm, and data direct reporting to the health supervision bureau make the equipment suitable for online water quality monitoring in water supply, swimming pools, sewage treatment, and aquaculture.
★★★★★
Multi Parameter Analyzer Features
Specifications
| Category | Specification |
|---|---|
| Measuring Parameters & Ranges | pH: 0.01~14.00pH; ORP: -1999~+1999mV; Free Chlorine: 0.01~20.0mg/L; Turbidity: 0~20NTU; Temperature: 0.1~100.0℃ |
| Measuring Principles | pH/ORP: Glass Electrode Method; Free Chlorine: Constant Voltage Method; Turbidity: 90° Scattered Light Method; Temperature: PT1000 Thermistor Method |
| Display Interface | 7″ TFT touch screen, LED backlight, 800×480 resolution |
| Power Supply | 100-240VAC or 9-36VDC (optional) |
| Protection Grade | IP54/IP55 |
| Overall Dimension | 750×600×250mm (H×W×D) |
| Output Interfaces | RS485/RS232 (Modbus RTU), AIP data interface; 4 sets of editable active relays |
| Wireless Remote | DTU optional, supporting 4G/WIFI/Ethernet |
| Data Storage | Stores 60 days of historical data, supports report export and historical curve viewing |
| Flow Regulation | Dual flow regulation; Free Chlorine: 18-30L/H, Turbidity: ≤18L/H |
Featured Applications
- Swimming Pool/Hot Spring: Swimming Pool, Water Park, Hot Spring Water Residual Chlorine pH、 Real time monitoring of turbidity, in compliance with hygiene standards;
- Wastewater treatment: monitoring the treatment process and effluent quality of municipal and industrial wastewater to ensure compliance with discharge standards;
- Aquaculture: Monitoring of water quality in fish ponds and breeding bases, precise regulation of water quality environment, and ensuring the survival of aquaculture organisms;
- Industrial cooling: Monitoring the quality of industrial cooling water to prevent water deterioration from affecting equipment operation;

Daily maintenance and precautions
Installation environment requirements
- Installation in a dry indoor environment, avoiding direct sunlight, strong electromagnetic field interference, and contact with highly corrosive gases;
- The ambient temperature is maintained at 0-50 ℃, the relative humidity is 10-90% RH, and there is no risk of severe shaking or falling;
- Reserve operating space at the installation location for later equipment debugging, maintenance, and pipeline operations.
Key daily maintenance points
- Electrode maintenance: pH/ORP electrodes should be cleaned with distilled water after use and soaked in a 3.3Mol/L potassium chloride solution when idle; The residual chlorine electrode should avoid breaking through the chlorine membrane, and the membrane head should be cleaned in a timely manner to prevent scaling. When the electrolyte is missing, it should be replenished in a timely manner;
- Sensor calibration: It is recommended to calibrate pH/ORP every 1-3 months, turbidity every 3 months, clean every 6 months, and shorten the cycle appropriately in scenarios with poor water quality; All parameters support convenient on-site calibration and are easy to operate;
- Equipment cleaning: Regularly wipe the surface of the cabinet and touch screen to keep the equipment clean; Check whether the pipeline connection is firm, prevent water leakage, and keep the overflow and sewage outlet unobstructed;
- Wiring inspection: Regularly check whether the wiring terminals are dry and free of dirt. If there is dirt, wipe and dry it with anhydrous alcohol to avoid poor contact;
- Data management: Regularly export historical data for archiving to avoid data loss; Regularly check the remote transmission module to ensure smooth network connectivity and normal data upload.
Technical Support
Top 6 Industrial Gas Analyzers Manufacturers 2025
6 Best Industrial Gas Detector Manufacturers in the World
9 Best Portable Carbon Dioxide Detector 2025 – Complete Buying Guide
Top 7 portable combustion analyzer Tested
The DO-PA100 portable fluorescence-based dissolved oxygen detector blocks interference, protects your eyes from the ink screen, and switches modes easily, making it ideal for scene monitoring. It can reliably monitor aquaculture 24/7, take short outdoor water quality inspections, and accurately monitor the environment. This simplifies operation and lowers maintenance costs.
We provide professional technical support and after-sales service, covering the entire process of installation guidance, calibration training, troubleshooting, etc. If you have any questions during selection and use, please feel free to contact us at any time.
Request a Quote
.png)