GRT1020CW is a professional plastic shell pH electrode specially designed for industrial online detection. It is equipped with antifouling PTFE connector and gel electrolyte to solve common problems such as blockage, poisoning and frequent maintenance, and ensure stable and reliable pH measurement in complex water environment.
★★★★★
Features
Specifications
| Parameter | Value |
|---|---|
| Measurement Range | 0-14 pH |
| Theoretical percentage slope | ≥95% |
| Zero potential | 7±0.5 |
| Temperature Range | 0-60℃ |
| Pressure Range | 0-0.4 M Pa |
| Internal Resistance | ≤ 250 MΩ at 25 degrees |
| Thread Size | 3/4 inch NPT |
| Service Life | 6 months (under normal use) |
Scope of Application
- The sensor can be used in conjunction with various acidity meters and is suitable for online detection of pure water;
- If the sensor is used normally for six months, its lifespan may be shortened due to harsh environments or improper maintenance.
Usage, maintenance, and precautions
- When measuring, it should be washed in distilled water (or deionized water) first, and the water should be absorbed dry with filter paper to prevent impurities from entering the measured liquid. The sensor bulb and liquid trap should be immersed in the measured liquid at the same time.
- When the sensor is not in use, it should be cleaned and covered with a protective cover soaked in 3.3 Mol/L KCL solution.
- Wastewater pH sensors should avoid long-term immersion in distilled water, protein solutions, and acidic chloride solutions, and prevent contact with organic silicon oils and fats.
- For sensors that have been used for a long time, their glass film may become semi transparent or have sediment attached. In this case, dilute hydrochloric acid can be used to wash and rinse with water.
- It is recommended that users clean the sensors once a week and cooperate with instrument calibration.
- When you fail to perform the calibration procedure on the sensor using the above methods for maintenance and upkeep, the sensor will no longer be able to recover its response. Please replace the electrode.
Applications
- Water Treatment: This is a great way to keep an eye on the pH of both clean water and sludge.
- Use in industry: It works well in harsh conditions where accurate, long-lasting pH values are needed.
- A good way to get accurate pH readings for studies is to test things in the lab.
Installation & Setup
- Check the sensitive bulb for any mechanical damage and make sure there are no air bubbles before using it.
- To make sure the installation is safe, use the right electrode protective sleeve or cable connector.
- Waterproofing the sensor’s connecting point correctly is important to keep water from getting in.
Important Notes
- The electrode must be rinsed with distilled or deionized water before measurement to avoid contamination from other substances.
- The electrode’s glass membrane may become opaque over time. If so, clean it with dilute hydrochloric acid to restore performance.
Sensor diagram illustration

- After opening the packaging, check whether there is mechanical damage to the sensitive bulb and electrode body of the electrode.
- Remove the sensor protective cover and check if there are any bubbles in the sensitive bulb of the electrode. If so, gently shake the electrode in a vertical plane to remove the bubbles.
- Select the corresponding sensor sleeve or cable head for installation according to different electrodes.
- The connection between the sensor wire and the electrode is not waterproof, please make sure to do a good job of waterproofing.
Related Products & Accessories
| Product Name | Function | Model |
| pH Controller | Single-channel online monitoring & control | BT-280 |
| Dual-Parameter Controller | pH/ORP simultaneous monitoring | A11PR |
| Immersion Bracket | Fixed installation for tanks/pools | PP-01 |
| Pipeline Adapter | Pipeline installation for process lines | SS316-03 |
| KCl Soaking Solution | Electrode storage & maintenance | KCL-3.3 |
Technical Support
5 Best carbon monoxide detector for garage
10 Best carbon monoxide monitor for industrial use 2025
7+ Best Recommended Carbon Monoxide Detector for Industrial Use
pH In Water: Importance & Buying Guide
6 Types of Industrial Oxygen Sensors
6 Best Industrial Gas Detector Manufacturers in the World
Choosing a suitable plastic shell pH sensor is crucial for achieving stable, accurate, and durable water quality measurements. Whether you are managing wastewater, aquaculture, or general industrial processes, our electrodes offer reliable performance, stronger durability, and exceptional value compared to traditional options.
If you are looking for a sensor that is easy to install, resistant to pollution, and can meet real-world challenges, grt1020cw is a reliable and wise choice for engineers and operators in multiple industries.
Request a Quote
.png)